Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 98447

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 03, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
GE Medical Systems, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Imactis CT-Navigation System, stereotaxic accessory for Computer Tomography (CT), systems, Model Numbers E85101LA and E85101LP

Z-1590-2026
Recall number
Z-1590-2026
Initiated
February 03, 2026
Classification
Class II
Status
Ongoing
Recalling firm
GE Medical Systems, LLC
Quantity
125 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential issue that can lead to discordance between the simulated needle trajectory/tip position and the actual needle trajectory/tip position.

Code information

HIBC: +B681J02000; E85101LA: IMB-22-013 IMB-24-013 IMB-23-006 IMB-24-001 IMB-24-016 IMB-24-018 IMB-24-007 IMB-21-001 IMB-23-004 IMB-24-015 IMB-22-014 IMB-22-011 IMB-21-008 IMB-22-010 IMB-23-003 IMB-22-002 IMB-16-005 IMB-16-002 IMB-17-007 IMB-21-003 IMB-16-003 IMB-18-006 IMB-16-006 IMB-19-014 IMB-23-007 IMB-19-011 IMB-24-011 IMB-15-001 IMB-24-009 IMB-20-005 IMB-19-009 IMB-15-003 IMB-18-007 IMB-17-011 IMB-18-003 IMB-19-003 IMB-17-006 IMB-20-004 IMB-21-005 IMB-21-006 IMB-20-012 IMB-22-004 IMB-17-008 IMB-20-022 IMB-19-019 IMB-18-004 IMB-17-012 IMB-19-008 IMB-15-004 IMB-16-004 IMB-17-005 IMB-18-002 IMB-18-005 IMB-18-010 IMB-19-005 IMB-15-002 IMB-17-009 IMB-17-010 IMB-24-012 IMB-17-002 IMB-20-016 IMB-22-007 IMB-24-002 IMB-20-010 IMB-22-005 IMB-22-017 IMB-24-021 IMB-19-002 IMB-24-017 IMB-20-021 IMB-19-007 IMB-24-014 IMB-24-004 IMB-24-020 IMB-19-013 IMB-20-009 IMB-24-006 IMB-20-008 IMB-24-023 IMB-24-022 IMB-18-009 IMB-19-001 IMB-19-004 IMB-19-006 IMB-19-010 IMB-19-016 IMB-22-001 IMB-22-012 IMB-22-016 IMB-22-018 IMB-17-003 IMB-24-005 IMB-16-007 IMB-22-015 IMB-24-003 IMB-20-002 IMB-20-003 IMB-20-017 IMB-20-018 IMB-20-019 IMB-21-002 IMB-21-004 IMB-22-003 IMB-22-008 IMB-22-009 IMB-20-011 IMB-20-020; E85101LP: IMB-25-005 IMB-21-007 IMB-25-007 IMB-25-008 IMB-20-001 IMB-20-013 IMB-25-012 IMB-25-009 IMB-25-006 IMB-21-009 IMB-25-013 IMB-25-010 IMB-19-012 IMB-25-001 IMB-25-004 IMB-24-024 IMB-25-023 IMB-25-003

Distribution pattern

Worldwide distribution - US Nationwide.

Field note

Send feedback

We'll only use this to respond to your feedback.