Open FDA event record

Event summary

Timeline bucket
February 18, 2026
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Baxter Healthcare Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software Version: Software version 3.2.0.1 Product Description: Welch Allyn Spot Vision Screener VS100, FDA product code name: Refractometer, ophthalmic One unit per package

Z-1573-2026
Recall number
Z-1573-2026
Initiated
February 18, 2026
Classification
Class II
Status
Ongoing
Quantity
59

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software issue

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.

Code information

Lot Code: " Product Code: HKO " Version or Model: 901029 " Catalog Number: VS100 " Company Name: WELCH ALLYN, INC. " Primary DI Number: 00732094214154 " Packaging DI Number: 00732094212228 " Device Description: VS100 UDI BASE UNIT " Software version: 3.2.0.1 " Serial Numbers: Refer to attached Affected Product Table

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.

Field note