DxC 700 AU, REF: B86444, B86446
- Recall number
- Z-1552-2026
- Initiated
- January 30, 2026
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Beckman Coulter Mishima K.K.
- Quantity
- 142
App-derived interpretation
software solution; may
Official device-enrichment evidence · Unavailable
No enrichment row joined through the supported exact identifiers. This is missing evidence, not evidence that no root cause exists.
Inspect official wording and provenance
Reason for recall
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.
Code information
REF/UDI-DI/Serial Numbers: Newer Instrument Installed with PC preinstalled with Microsoft SQL Server 2022 Express: B86444/14987666542545/2025060063, 2025060073, 2025060074, 2025060075, 2025060079, 2025060085, 2025060086, 2025060087, 2025060091, 2025060098, 2025060099, 2025060124, 2025060125, 2025060126, 2025060127, 2025060128, 2025060129, 2025070139, 2025070156, 2025070159, 2025070160, 2025070161, 2025070164, 2025070167, 2025070168, 2025070169, 2025070178, 2025070180, 2025070181, 2025070182, 2025070184, 2025070186, 2025070191, 2025070192, 2025070193, 2025070194, 2025070204, 2025080208, 2025080217, 2025080219, 2025080236, 2025080237, 2025080238, 2025080241, 2025080242, 2025080246, 2025080247, 2025080250, 2025090266, 2025090267, 2025090270, 2025090271, 2025090273, 2025090274, 2025090275, 2025090281, 2025090283, 2025090285, 2025090286, 2025090287, 2025090289, 2025090290, 2025090291, 2025090292, 2025090293, 2025090294, 2025090295, 2025090296, 2025090297, 2025090298, 2025090299, 2025090301, 2025090302, 2025090303, 2025090304, 2025090306, 2025100370, 2025100371, 2025100372, 2025100373, 2025100421, 2025100422, 2025100423, 2025100424, 2025100425, 2025100426, 2025100427, 2025110428, 2025110429, 2025110437, 2025110442, 2025110445, 2025110452, 2025110454, 2025110456, 2025110457, 2025110458, 2025110462, 2025120464, 2025120468, 2025120469, 2025120471, 2025120475, 2025120477, 2025120478, 2025120479, 2025120480, 2025120482, 2025120483, 2025120484, 2025120485, 2025120486, 2025120487, 2025120488, 2025120489, 2025120490, 2025120492, 2025120493, 2025120494, 2025120495, 2025120496. B86446/14987666542514/2025090277, 2025090284. Older Instrument Replaced with PC preinstalled with Microsoft SQL Server 2022 Express: B86444/14987666542545/2017050064, 2017100174, 2018030312, 2018040352, 2018100561, 2018110599, 2019010712, 2019010713, 2019111170, 2021021852, 2021102318, 2021122458, 2022042631, 2022052705, 2022062775, 2023063701, 2024064408, 2024064419, 2024064440
Distribution pattern
US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.