Event 98111

13 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 03, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Philips North America

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

13 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 13

Evolution Upgrade 1.5T. Product Codes (REF): (1) 782148, (2) 782116. MR systems with SW version R11.1 and R12.1.

Z-1204-2026
Recall number
Z-1204-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
3 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 782148. UDI-DI: 00884838108714. Serial Numbers: 70769, 70983. (2) Product Code (REF): 782116. UDI-DI: 00884838099722. Serial Number: 84741.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 2 of 13

Evolution Upgrade 3.0T. Product Code (REF): 782143. MR systems with SW version R11.1 and R12.1.

Z-1205-2026
Recall number
Z-1205-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
1 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. Product Code (REF): 782143. UDI-DI: 00884838108660. Serial Numbers: 42225.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 3 of 13

Ingenia 1.5T. Product Code (REF): (1) 781341, (2) 781396. MR systems with SW version R11.1 and R12.1.

Z-1206-2026
Recall number
Z-1206-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
10 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781341; UDI-DI: 00884838055322; Serial Numbers: 70311, 84560, 84403, 84467, 84462, 70042, 84433. (2) Product Code (REF): 781396; UDI-DI: 00884838009820; Serial Numbers: 41356, 41354, 41370.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 4 of 13

Ingenia 3.0T. Product Code (REF): (1) 781342, (2) 781377. MR systems with SW version R11.1 and R12.1.

Z-1207-2026
Recall number
Z-1207-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
24 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781342; UDI-DI: 00884838055339; Serial Numbers: 71607, 85216, 71087, 71619, 71462, 71002, 71885, 71841, 71378, 71269, 85126, 71279, 71088, 71185, 71532, 71344, 71756, 71336, 71459. (2) Product Code (REF): 781377; UDI-DI: 00884838009813; Serial Numbers: 42305, 42299, 42036, 42376, 42321.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 5 of 13

Ingenia 3.0T CX. Product Code (REF): 781271. MR systems with SW version R11.1 and R12.1.

Z-1208-2026
Recall number
Z-1208-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
6 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. Product Code (REF): 781271. UDI-DI: 00884838068452. Serial Numbers: 78017, 78112, 78217, 78123, 78260, 78239.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 6 of 13

Ingenia Ambition S. Product Code (REF): 782108. MR systems with SW version R11.1 and R12.1.

Z-1209-2026
Recall number
Z-1209-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
1 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. Product Code (REF): 782108. UDI-DI: 00884838098343. Serial Numbers: 48669.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 7 of 13

Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.

Z-1210-2026
Recall number
Z-1210-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
21 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781356; UDI-DI: 00884838090040; Serial Numbers: 47206, 47038, 47068, 47260, 47294, 47029, 47189. (2) Product Code (REF): 782109; UDI-DI: 00884838098350; Serial Numbers: 47849, 47634, 47534, 47726, 47760, 47706, 47789, 47642, 47681, 47761, 47771, 47767, 47862. (3) Product Code (REF): 782138; UDI-DI: 00884838108622; Serial Numbers: 49057.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 8 of 13

Ingenia Elition S. Product Code (REF): (1) 781357, (2) 782106, (3)782137. MR systems with SW version R11.1 and R12.1.

Z-1211-2026
Recall number
Z-1211-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
5 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781357; UDI-DI: 00884838088108; Serial Numbers: 46002. (2) Product Code (REF): 782106; UDI-DI: 00884838098329; Serial Numbers: 46243, 46292. (3) Product Code (REF): 782137; UDI-DI: 00884838108615; Serial Numbers: 64034, 64045.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 9 of 13

Ingenia Elition X. Product Code (REF): (1) 781358, (2) 782107, (3) 782136. MR systems with SW version R11.1 and R12.1.

Z-1212-2026
Recall number
Z-1212-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
31 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781358; UDI-DI: 00884838088115; Serial Numbers: 45414, 45245, 45090, 45049, 45034, 45250, 45075, 45290, 45320, 45011, 45191, 45277, 45278, 45028, 45147. (2) Product Code (REF): 782107; UDI-DI: 00884838098336; Serial Numbers: 62031, 45533, 62051, 62182, 62024, 62047, 62048, 62052, 45519, 62181, 62081, 62153, 62088, 45589, 62073. (3) Product Code (REF): 782136; UDI-DI: 00884838108608; Serial Numbers: 28503.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 10 of 13

MR 7700. Product Code (REF): (1) 782120, (2) 782153. MR systems with SW version R11.1 and R12.1.

Z-1213-2026
Recall number
Z-1213-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
9 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 782120; UDI-DI: 00884838104112; Serial Numbers: 65019, 65060, 65066, 65065, 65051, 65038, 65005. (2) Product Code (REF): 782153; UDI-DI: 00884838112858; Serial Numbers: 68010, 68017.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 11 of 13

Upgrade to MR 7700. Product Code (REF): 782130. MR systems with SW version R11.1 and R12.1.

Z-1214-2026
Recall number
Z-1214-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
6 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. Product Code (REF): 782130. UDI-DI: 00884838104402. Serial Numbers: 42030, 85644, 42163, 42151, 45236, 85643.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 12 of 13

SmartPath to dStream for 3.0T. Product Code (REF): 782145. MR systems with SW version R11.1 and R12.1.

Z-1215-2026
Recall number
Z-1215-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
1 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. Product Code (REF): 782145. UDI-DI: 00884838108684. Serial Numbers: 17385.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.

device · product 13 of 13

SmartPath to dStream for XR and 3.0T. Product Code (REF): 781270. MR systems with SW version R11.1 and R12.1.

Z-1216-2026
Recall number
Z-1216-2026
Initiated
December 03, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Philips North America
Quantity
3 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

Code information

MR systems with SW version R11.1 and R12.1. Product Code (REF): 781270. UDI-DI: 00884838095083. Serial Numbers: 24066, 38113, 17356.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, IL, MA, ME, MI, MN, MS, NE, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and the countries of Belgium, Canada, Denmark, France, Germany, Greece, India, Israel, Japan, Kenya, Netherlands, South Korea, Spain, Sweden, Taiwan, Thailand, United Kingdom.