device · product 1 of 1
GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use.
- Recall number
- Z-0036-2026
- Initiated
- August 26, 2025
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- W L Gore & Associates, Inc.
- Quantity
- 52,306
App-derived interpretation
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
Code information
REF/UDI-DI: ECH050020J/04993024010758, ECH050020W/00733132635450, ECH050050J/04993024010772, ECH050050W/00733132635474, ECH060010A/00733132622573, ECH060020A/00733132622597, ECH060020J/04993024009967, ECH060020W/00733132634729, ECH060040/00733132605347, ECH060040A/00733132605354, ECH060040W/00733132619719, ECH060050A/00733132626960, ECH060050J/04993024009981, ECH460045A/00733132634811, ECH460045J/04993024010789, ECH470045/00733132626977, ECH470045A/00733132634811. Product with a serial number that includes and falls within the range of 6597081PP001 and 9853822PP024. Instructions for Use MD183888 Revision 4
Distribution pattern
US: OR, AL, OH, MD, VA, CA, FL, NC, OK, MA, MO, AZ, TX, IL, GA, PA, NH, HI, MN, MS, KY, ID, LA, NJ, TN, DE, IN, MI, WI, NY, DC, SC, CT, NE, IA, WA, AR, NV, WV, AK, UT, ND, KS, CO, ME, MT, NM, SD, PR, WY, RI. OUS: PANAMA, CANADA, CHINA, TAIWAN, GERMANY, SLOVENIA, NORWAY, UNITED KINGDOM, ITALY, SPAIN, FINLAND, NETHERLANDS, SWEDEN, FRANCE, UNITED ARAB EMIRATES, DENMARK, PORTUGAL, AUSTRIA, SWITZERLAND, BELGIUM, TURKIYE, SLOVAK REPUBLIC, CYPRUS, SAUDI ARABIA, SOUTH AFRICA, LUXEMBOURG, FRENCH GUIANA, ANDORRA, IRELAND, NEW CALEDONIA, POLAND, ISRAEL, GREECE, GUADELOUPE, CROATIA, REUNION, BAHRAIN, HUNGARY, JAPAN