Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 97403

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 05, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Tandem Diabetes Care, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

Z-0232-2026
Recall number
Z-0232-2026
Initiated
August 05, 2025
Classification
Class II
Status
Ongoing
Quantity
122,958

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software defect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

Code information

t:slim X2, UDI-DI: 00389152407012, 00389152407111, 00389152407210, 00389152407319 with software versions 7.9.0.1 and 7.10.1, User Guide versions not including and prior to: AW-1014562 Revision D; Tandem Mobi, UDI-DI: 00389152480114, 00389152481111, 00389152481319 with software version 7.9.0.1, User Guide versions not including and prior to: AW-1014938 Revision C

Distribution pattern

US nationwide distribution including Puerto Rico.

Field note

Send feedback

We'll only use this to respond to your feedback.