Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96485

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 20, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Channel Medsystems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Brand Name: Cerene Cryotherapy Device Product Name: DEVICE, THERMAL ABLATION, ENDOMETRIAL Model/Catalog Number: FGS-7000 Product Description: The Cerene Cryotherapy Device is an endometrial ablation device that uses nitrous oxide (N2O) to freeze and ablate the endometrium to reduce future menstrual bleeding. The device is intended for use by healthcare professionals who have received appropriate training and are familiar with the principles, clinical applications, complications, side effects, and hazards commonly associated with endometrial ablation. The Cerene Device is single use, disposable, and provided sterile. The average procedure time is 7 minutes, with 2.5 minutes of active cryoablation. Ablation is achieved throughout the uterine cavity through the use of cryothermic energy. The cryothermic energy is provided by a liquid-to-gas phase change of N2O. During the 2.5-minute treatment cycle, liquid N2O (originating from a small Cylinder located in the device handle) flows through a delivery line and into an inflow line with multiple jets. This liquid N2O is infused into an ultra-thin polyurethane Liner, where it converts into gas. The gaseous N2O is exhausted through the Exhaust Hose exiting the bottom of the handle. Component: N/A

Z-1584-2025
Recall number
Z-1584-2025
Initiated
March 20, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Channel Medsystems, Inc.
Quantity
281

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to evaporator gasket not sealing which can cause liquid nitrous oxide to drip/leak from the bottom of the device handle during treatment or after treatment during venting of the device

Code information

Lot Code: Model No: FGS-7000 UDI-DI: 00850008595035 Lot Numbers: 108638953, 108645585, 108645586, 108645587, 108866674, 108866675, 109493291, 109493292 and 109895874

Distribution pattern

U.S Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IL, IN, KY, LA, MD, MO, NC, NY, OH, PA, PR, RI, TN, TX.

Field note

Send feedback