Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96197

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 23, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
LUMENIS, LTD.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.

Z-1290-2025
Recall number
Z-1290-2025
Initiated
January 23, 2025
Classification
Class II
Status
Ongoing
Recalling firm
LUMENIS, LTD.
Quantity
3 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

Code information

Software version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.

Distribution pattern

Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.

Field note

Send feedback

We'll only use this to respond to your feedback.