Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 96153

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 15, 2025
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Elekta, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

MOSAIQ Oncology Information System

Z-1162-2025
Recall number
Z-1162-2025
Initiated
January 15, 2025
Classification
Class II
Status
Ongoing
Recalling firm
Elekta, Inc.
Quantity
64 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software may be incorrect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Image offsets calculated by oncology Information System software may be incorrect for Linear Accelerator (LINACs) not characterized at IEC61217 scale.

Code information

Software versions /UDI: 3.1.2.0 (01)07340201500026 3.1.3.0 (01)07340201500026; 3.1.3.1 (01)07340201500026; 3.1.3.2 (01)07340201500026; 3.1.3.3 (01)07340201500026; 3.1.3.4 (01)07340201500026; 3.2.1.0 (01)07340201500071; 3.2.1.1 (01)07340201500071; 3.2.1.2 (01)07340201500071; 3.2.1.3 (01)07340201500071; 3.2.1.4 (01)07340201500071

Distribution pattern

US: PA, AL, FL, IA UT OUS: Austria Belgium Bolivia Bosnia and Herzegowina Brazil Canada France Germany Hong Kong Italy Japan Korea, Republic of Malta Netherlands Panama South Africa Spain Taiwan Thailand Turkey Venezuela

Field note

Send feedback