Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94814

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 22, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Waldemar Link GmbH & Co. KG (Mfg Site)

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Endo Model M Tibial Components Modular Knee Prosthesis System. Product codes 15-2814/01, 15-2814/02, 15-2814/03, 15-2814/04, 15-2815/11, 15-2815/12, 15-2816/11, 15-2816/12, 15-2817/11, 15-2817/12, 15-2818/11, 15-2818/12, 15-2834/01, 15-2834/02, 15-2834/03, 15-2834/04, 15-2835/11, 15-2835/12, 15-2836/11, 15-2836/12, 15-2837/11, 15-2837/12, 15-2838/11, 15-2838/12, 15-3815/11, 15-3815/12, 15-3816/11, 15-3816/12, 15-3817/11, 15-3817/12, 15-3818/11, 15-3818/12, 15-8521/05, 15-8521/07, 15-8521/09, 15-8521/11, 15-8521/15, 15-8521/25, 15-8521/27, 15-8521/29, 15-8521/31, 15-8521/33

Z-2410-2024
Recall number
Z-2410-2024
Initiated
May 22, 2024
Classification
Class II
Status
Ongoing
Quantity
552 units; 7351 units OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique

Code information

Product UDI-DI: 1 15-2814/01 04026575316243 2 15-2814/02 04026575316281 3 15-2814/03 04026575316298 4 15-2814/04 04026575316304 5 15-2815/11 04026575316007 6 15-2815/12 04026575316014 7 15-2816/11 04026575316021 8 15-2816/12 04026575316038 9 15-2817/11 04026575316052 10 15-2817/12 04026575316069 11 15-2818/11 04026575316076 12 15-2818/12 04026575316083 13 15-2834/01 04026575316731 14 15-2834/02 04026575316748 15 15-2834/03 04026575316755 16 15-2834/04 04026575316762 17 15-2835/11 04026575316557 18 15-2835/12 04026575316564 19 15-2836/11 04026575316571 20 15-2836/12 04026575316588 21 15-2837/11 04026575316595 22 15-2837/12 04026575316601 23 15-2838/11 04026575316618 24 15-2838/12 04026575316625 25 15-3815/11 04026575034727 26 15-3815/12 04026575034734 27 15-3816/11 04026575034741 28 15-3816/12 04026575034758 29 15-3817/11 04026575164042 30 15-3817/12 04026575164059 31 15-3818/11 04026575164028 32 15-3818/12 04026575164035 33 15-8521/05 04026575383207 34 15-8521/07 04026575383221 35 15-8521/09 04026575383238 36 15-8521/11 04026575383245 37 15-8521/13 04026575383269 38 15-8521/15 04026575383283 39 15-8521/25 04026575386406 40 15-8521/27 04026575386413 41 15-8521/29 04026575386420 42 15-8521/31 04026575386437 43 15-8521/33 04026575386444 44 15-8521/35 04026575386451 All Product lots manufactured since 01-Jun-2022.

Distribution pattern

Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slovakia Slovenia Spain Sweden Switzerland Thailand T¿rkiye Ukraine United Arab Emirates United Kingdom Uruguay Vietnam

device · product 2 of 2

Endo Model SL Tibial Components 16-2817/02 Tibial Component, Small, W: 60mm 16-2817/05 Tibial Component, Medium, W: 65mm 16-2817/07 Tibial Component, Large, W: 75mm 16-2817/32 Tibial Component Modular, Small, W: 60mm 16-2817/35 Tibial Component Modular, Medium, W: 65mm 16-2817/37 Tibial Component Modular, Large, W: 75mm

Z-2411-2024
Recall number
Z-2411-2024
Initiated
May 22, 2024
Classification
Class II
Status
Ongoing
Quantity
114 units US; 2351 units OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Blind screws of the modular tibial component cannot be loosened intraoperatively,prolongation surgery due to intraoperatively change in procedure, probably to cementing technique

Code information

Product UDI-DI 16-2817/02 04026575359202; 16-2817/05 04026575359219; 16-2817/07 04026575359226; 16-2817/32 04026575370870; 16-2817/35 04026575370887; 16-2817/37 04026575370894. All product lots manufactured since 01-Jun-2022.

Distribution pattern

Nationwide Foreign: Algeria Argentina Australia Austria Belarus Belgium Brazil Bulgaria Canada China Colombia Czechia Denmark Ecuador Estonia Finland France Germany Greece Hungary India Indonesia Israel Italy Kenya Lithuania Luxembourg Mexico Netherlands Norway Peru Poland Romania Saudi Arabia Slovakia Slovenia Spain Sweden Switzerland Thailand T¿rkiye Ukraine United Arab Emirates United Kingdom Uruguay Vietnam

Field note

Send feedback

We'll only use this to respond to your feedback.