device · product 1 of 1
MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)
- Recall number
- Z-2041-2024
- Initiated
- May 10, 2024
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Canon Medical System, USA, INC.
- Quantity
- 153
App-derived interpretation
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.
Code information
Model/UDI-DI/Serial Numbers: MRT-1550 (MEXL-1550)/04987670103360/POA21X2002, S5A1982001, S5B22Z2041, S5B2172018, S2C2252024, S5A20Z2009, S2C21Y2021, S2B2052010, S5B2272031, POA2262005, S5B2292036, S5B2272029, S5B2282034, S5B22Y2040, S5B2292035, S6B2252008, S5A20Z2010, POA2252004, POA2182001, S1C2152071, S5B2352047, S2C2252023, S5B2122013, S6B2262010, S2C2192020, S6B2292013, S5B2132014, S6B2322018, S5B2292038, S6B2312016, S2B2052011, S2C2162018, S4A1992013, S3A19Z2013, S3A1982009, S5B2182021, S6B2322017, S5A2092008, S6B2362019, S5B22Z2042, S6B2222005, S2B2042009, S2B20Z2015, S5B2162017, S6B2262011, S5B2272030, S6B22Z2014, S5B2282033, S8B2272009, S5A2062007, S2B2072012, S2B2072013, S6B2222006, S6B2272012, S2B2072014, S6A20Z2001, POA2272006, S5A19Z2003, S1C2142070, POA2212003, S2C2142017, S5B2362048, S2A1962005, S6B2182004, S5B2292037, S6B2252007, S5B2282032, S5B2142015, S5B2382050, S2A18Z2003, S6B2112002, S6B2382021, S1B2042051, S8B2212007, S1B2062057, S1A1922022, S2A18Z2001, S1B2092067, S1C2242080, S5B2172019, S6B2252009, S6B2372020, S2A18Z2002, S2A1942004, S2C2212022, S2C2272025, S6B2392022, S2C2162019, S2C2132016, S4B2062020, S1C2262081, PDA21Y2001, S6B22Z2015, S5B22Z2043, S5B2242027, MRT-2020 (MEXL-1520)/04987670101632/S1F18X2580, S1B1492068, S1D17Y2461, S1F2052753, S1C1722390, S1C1682336, S1E1842517, S1E1842527, S1C1672334, MRT-1504 (MEXL-1504)/04987670102400/U5B15Z2001, MRT-3010 (MEXL-3010)/04987670101649/A5D14X2032, A5C12Y2017, A5D13Y2025, A5C1412028, S7A1642006, A5D13Y2026, A5C1212009, A5C12X2016, A5G1692038, A5G1642036, A5F1562034, A5B11Y2005, A5C1262013, A5E1522033, A5G1692039, A5D14X2031, A5B1192003, A5B11Y2006, A5D1312019, A5D1362021, A5C1252012, A5C1292015, A5D13X2024, A5C1422029, A5B11Z2007, A5C1212008, MRT-3020 (MEXL-3020)/04987670102615/D5A2082005, D5A2092007, D5B2152013, S5B17X2034, D7B2182017, 5DB2152002, S5B18X2046, S5A17X2033, D5A2062003, S5A1742021, D5B2212019, D5B2192016, S5A1662001, S5C18Y2048, S5A1732019, S5A1732018, S5A17Y2035, S7B1892004, S5B1842043, D7B2152015, S5C1972054, 5DB2172003
Distribution pattern
US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR