Timeline bucket
November 15, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Canon Medical System, USA, INC.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

The device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in an interventional setting. The system consists of a C-arm which is equipped with an X-ray tube, beam limiter and X-ray receptor, X-ray controller, computers with system and processing software, and a patient radiographic table.

Z-1154-2024
Recall number
Z-1154-2024
Initiated
November 15, 2023
Classification
Class II
Status
Ongoing
Quantity
1

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Code information

UDI-DI: 04987670100192, Serial Number: B1E22X2005

Distribution pattern

US: CA, LA

device · product 2 of 2

The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.

Z-1155-2024
Recall number
Z-1155-2024
Initiated
November 15, 2023
Classification
Class II
Status
Ongoing
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

Code information

UDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017

Distribution pattern

US: CA, LA