device · product 1 of 1
ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844
- Recall number
- Z-1007-2024
- Initiated
- December 15, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Globus Medical, Inc.
- Quantity
- 570
App-derived interpretation
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Code information
UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FSA286AA FSA263DA FSA263EA FSA263CA FSA263AA FSA263BA FSA160AA FSA160BA FSA105BA FSA105AA FSA053AA FSY319AA FSY279BA FSY279AA FSY272AA FSY230AA FSY098AA FSY098CA FSY098BA FSY048AA FSX093AA FSX205AA FSX141BA FSX141CA FSX141AA FSX093AA FSX063AA FSX041AA FSX041BA FSW142AA FSW094AA
Distribution pattern
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Czech Republic, Germany, India, Indonesia, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Saudi Arabia, Singapore, Spain, United Arab Emirates, United Kingdom.