Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 93738

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 28, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
XTANT Medical Holdings, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Cortera Spinal Fixation System, REF: 150100-85050. Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length.

Z-0846-2024
Recall number
Z-0846-2024
Initiated
November 28, 2023
Classification
Class II
Status
Ongoing
Quantity
6

App-derived interpretation

Labeling or packaging reason.labeling_packaging · v1.0.0
labeling is incorrect

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Spinal fixation screws labeled as 8.5 mm diameter, 50 mm length were manufactured using an incorrect screw shank component, and so the labeling is incorrect because the affected screws are in fact 5.5 diameter, 45 mm length screws, which if used may lead to the screw not engaging.

Code information

UDI-DI: 00810076564005, Lot: AP23000143

Distribution pattern

US distribution in MI.

Field note

Send feedback

We'll only use this to respond to your feedback.