device · product 1 of 4
NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176
- Recall number
- Z-0453-2024
- Initiated
- September 13, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Thor Photomedicine Ltd
- Quantity
- 19
App-derived interpretation
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
Code information
Serial numbers: 2003, 2004, 2005, 2006, 2007, 2009, 2010, 2011, 2017, 2018, 2015, 2020, 2019, 2001, 2021, 2028, 2033, 2036, 2030, 2029.
Distribution pattern
Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia. OUS (foreign) to countries of: Australia, Canada, Netherlands, New Zealand, Singapore, United Kingdom.