Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92886

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 21, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Ambu Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Ambu aView 2 A Monitor-The Ambu aView 2 Advance is intended to display live imaging data from compatible Ambu visualization devices, portable endoscopy system. Catalog No: 405011000

Z-2628-2023
Recall number
Z-2628-2023
Initiated
July 21, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Ambu Inc.
Quantity
2689 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Multiple impacts, for example from being dropped, can cause to the Ambu aView 2 Advance to short-circuit and under special circumstances make the lithium-ion battery in the device combust leading to smoke and flames

Code information

UDI: 5707480145089 Version no.: 054 thru 056; 058 thru 059 Product manufactured prior to Oct 29, 2020

Distribution pattern

Nationwide

Field note

Send feedback

We'll only use this to respond to your feedback.