Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92576

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 20, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Ringed DxI Reaction Vessels (RVs)

Z-2237-2023
Recall number
Z-2237-2023
Initiated
June 20, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Beckman Coulter, Inc.
Quantity
213 cases of 10,000 RVs (2,130,000 RVs)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.

Code information

UDI: 15099590229979; Catalog Number: 386167; Lot Numbers: L17855111, L17798598, L18046334

Distribution pattern

Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.

Field note

Send feedback