device · product 1 of 1
Ringed DxI Reaction Vessels (RVs)
- Recall number
- Z-2237-2023
- Initiated
- June 20, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 213 cases of 10,000 RVs (2,130,000 RVs)
App-derived interpretation
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
Code information
UDI: 15099590229979; Catalog Number: 386167; Lot Numbers: L17855111, L17798598, L18046334
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, MA, TX, NJ, MI, GA, NC, FL, WV and the country of Qatar.