device · product 1 of 1
Covidien Cytosponge Cell Collection Device, REF: CYTO-201; and Covidien Cytosponge Cell Collection Kit, REF: CYTO-KIT-USM
- Recall number
- Z-2282-2023
- Initiated
- June 02, 2023
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Covidien, LLC
- Quantity
- 6902
App-derived interpretation
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
Code information
REF/UDI-DI/Lot: CYTO-201/20884521767789/22E1619JZX, 22F0680JZX, 22F0950JZX, 22G0007JZX, 22G0008JZX, 22G0723JZX, 22G0724JZX, 22H0017JZX, 22H0018JZX, 22H0019JZX, 22H0820JZX, 22H0821JZX, 22H1158JZX, 22H1159JZX, 22H1161JZX, 22J0162JZX, 22J0163JZX, 22J0164JZX, 22J1030JZX, 22K0966JZX, 22K0967JZX, 23B0301JZX, 23B0303JZX; CYTO-KIT-USM/10884521546257/22G0635JZX, 22E1624JZX, 22E1625JZX.
Distribution pattern
Worldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the countries of Canada, Netherlands, Poland, United Kingdom.