device · product 1 of 1
Integre Pro Yellow LP561, Integre Pro Yellow LP6RY, Integre Pro Scan Green LP6G, Integre Pro Green LP5532 Spare, Optics Bench V2, L2G/LP6G
57 systems in total (6 units shipped to United States)
App-derived interpretation
Unknownreason.no_named_rule · v1.0.0
When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.
These labels are deterministic app interpretations, not FDA categories.
When using a Laser-Indirect-Ophthalmoscope (LIO) as the delivery device, a scattered laser beam inside the laser cavity system may unintentionally exit through the Slit Lamp output port (SDS) and get emitted from the Objective Lens when the laser is fired.