Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 92460

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 13, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Wright Medical Technology, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Stryker Infinity Alignment Frame Distal Sub-Assembly, Catalog #33600020, non-sterile.

Z-2105-2023
Recall number
Z-2105-2023
Initiated
June 13, 2023
Classification
Class II
Status
Ongoing
Quantity
106 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.

Code information

Lot numbers 2656815 and 2762126, UDI-DI 00889797003926.

Distribution pattern

There was no U.S. distribution, including government and military. Foreign distribution was made to Australia, Belgium, Canada, Chile, China, Colombia, France, Germany, Italy, Mexico, Netherlands, Spain, Sweden, Switzerland, Taiwan, and the United Kingdom.

Field note

Send feedback

We'll only use this to respond to your feedback.