Event 92290

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 01, 2023
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Skytron, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

GS70 Salus Surgical Table, Product Code 18-070-70

Z-2469-2023
Recall number
Z-2469-2023
Initiated
August 01, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Skytron, LLC
Quantity
110 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Code information

Serial Numbers: SUC-20207-0006 SUC-20208-0025 SUC-20209-0044 SUC-20212-0063 SUC-20228-0082 SUC-2022Y-0101 SUC-20207-0007 SUC-20208-0026 SUC-20209-0045 SUC-20212-0064 SUC-20228-0083 SUC-2022Y-0102 SUC-20207-0008 SUC-20208-0027 SUC-2020X-0046 SUC-20212-0065 SUC-20228-0084 SUC-2022Y-0103 SUC-20207-0009 SUC-20208-0028 SUC-2020X-0047 SUC-20226-0066 SUC-20228-0085 SUC-2022Y-0104 SUC-20207-0010 SUC-20208-0029 SUC-2020X-0048 SUC-20226-0067 SUC-20228-0086 SUC-2022Y-0105 SUC-20207-0011 SUC-20209-0030 SUC-2020X-0049 SUC-20226-0068 SUC-20228-0087 SUC-20231-0106 SUC-20207-0012 SUC-20209-0031 SUC-2020X-0050 SUC-20226-0069 SUC-20228-0088 SUC-20231-0107 SUC-20207-0013 SUC-20209-0032 SUC-2020X-0051 SUC-20226-0070 SUC-20228-0089 SUC-20231-0108 SUC-20208-0014 SUC-20209-0033 SUC-2020X-0052 SUC-20227-0071 SUC-20228-0090 SUC-20231-0109 SUC-20208-0015 SUC-20209-0034 SUC-2020X-0053 SUC-20227-0072 SUC-20229-0091 SUC-20231-0110 SUC-20208-0016 SUC-20209-0035 SUC-2020X-0054 SUC-20227-0073 SUC-20229-0092 SUC-20231-0111 SUC-20208-0017 SUC-20209-0036 SUC-2020X-0055 SUC-20227-0074 SUC-20229-0093 SUC-20231-0112 SUC-20208-0018 SUC-20209-0037 SUC-2020Y-0056 SUC-20227-0075 SUC-20229-0094 SUC-20231-0113 SUC-20208-0019 SUC-20209-0038 SUC-2020Y-0057 SUC-20228-0076 SUC-20229-0095 SUC-20231-0114 SUC-20208-0020 SUC-20209-0039 SUC-2020Y-0058 SUC-20228-0077 SUC-2022X-0096 SUC-20231-0115 SUC-20208-0021 SUC-20209-0040 SUC-2020Y-0059 SUC-20228-0078 SUC-2022X-0097 SUC-20208-0022 SUC-20209-0041 SUC-2020Y-0060 SUC-20228-0079 SUC-2022X-0098 SUC-20208-0023 SUC-20209-0042 SUC-20212-0061 SUC-20228-0080 SUC-2022X-0099 SUC-20208-0024 SUC-20209-0043 SUC-20212-0062 SUC-20228-0081 SUC-2022X-0100

Distribution pattern

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.

device · product 2 of 2

GS70H Salus Surgical Table, Product Code 18-071-70

Z-2470-2023
Recall number
Z-2470-2023
Initiated
August 01, 2023
Classification
Class II
Status
Ongoing
Recalling firm
Skytron, LLC
Quantity
24 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Code information

Serial Numbers:SUC-20207-0002 SUC-20208-0006 SUC-20209-0010 SUC-2020X-0014 SUC-2022Z-0018 SUC-20233-0022 SUC-20207-0003 SUC-20208-0007 SUC-20209-0011 SUC-2020X-0015 SUC-2022Z-0019 SUC-20233-0023 SUC-20208-0004 SUC-20209-0008 SUC-2020X-0012 SUC-2020X-0016 SUC-2022Z-0020 SUC-20233-0024 SUC-20208-0005 SUC-20209-0009 SUC-2020X-0013 SUC-2022Z-0017 SUC-2022Z-0021 SUC-20233-0025

Distribution pattern

Worldwide - US Nationwide distribution in the states of AL, AR, CA, CT, FL, GA, IA, IL, IN, MA, MI, MO, PA, TN, TX, UT, WA, WI and the countries of Australia, New Zealand.