Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91639

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 12, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Healthcare Diagnostics Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

epoc Host 2 as follows: Product SMN epoc Host 10736387 epoc Host 2, Canada 10736388 epoc Host 2, China 10736389 epoc Host 2, Japan 10736390 epoc Host 2, Refurbished 10736393 epoc Host 2, Canada, Refurbished 10736394 epoc Host 2, Japanese, Refurbished 10736395 epoc Host 2, Chinese, Refurbished 10736433 epoc Host 2, US (MC55X) 11413524 epoc Host 2, ROW (MC55X) 11413528 epoc Host 2, India (MC55X) 11413541 epoc Host 2, China (MC55X) 11413542 epoc Host 2, Japan (MC55X) 11413543 epoc Host 2 (MC55X) - US - Refurbished 11413601 "epoc Host 2 (MC55X) - ROW -Refurbished" 11413602 epoc Host 2 (MC55X) - India - Refurbished 11413603 epoc Host 2 (MC55X) - Japan - Refurbished 11413604

Z-1214-2023
Recall number
Z-1214-2023
Initiated
December 12, 2022
Classification
Class II
Status
Ongoing
Quantity
22,156 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

Code information

Host Software Version /Sensor Configuration: 3.37.3/41.1 Product Description Siemens Material Number (SMN) Unique Device Identification (UDI-DI) epoc Host 2 10736387 00809708052898; epoc Host 2, Canada 10736388 00809708069117; epoc Host 2, China 10736389 00809708097219; epoc Host 2, Japan 10736390 00809708075576; epoc Host 2, Refurbished 10736393 00809708053901; epoc Host 2, Canada, Refurbished 10736394 00809708069124; epoc Host 2, Japanese, Refurbished 10736395 00809708075583; epoc Host 2, Chinese, Refurbished 10736433 00809708097226; epoc Host 2, US (MC55X) 11413524 00630414606460; epoc Host 2, ROW (MC55X) 11413528 00630414606378; epoc Host 2, India (MC55X) 11413541 00630414608617; epoc Host 2, China (MC55X) 11413542 00630414608624; epoc Host 2, Japan (MC55X) 11413543 00630414608532; epoc Host 2 (MC55X) - US - Refurbished 11413601 00630414615004; "epoc Host 2 (MC55X) - ROW - Refurbished" 11413602 00630414614885; epoc Host 2 (MC55X) - India - Refurbished 1413603 00630414614762; epoc Host 2 (MC55X) - Japan - Refurbished 11413604 00630414614892.

Distribution pattern

Nationwide Foreign: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kuwait Laos Latvia Lithuania Luxembourg Malaysia Mali Martinique Mexico Myanmar Netherlands New Caledonia New Zealand Norway Papua Nw Guinea Peru Philippines Poland Portugal Qatar Republic Korea Reunion Romania Russian Fed. San Marino Singapore Slovakia Slovenia Solomon Islands South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey U.A.E. Ukraine United Kingdom Uruguay Vietnam

device · product 2 of 2

epoc Host 2 as follows: Product SMN epoc NXS Host, US 11413475 epoc NXS Host, EU 11413497 epoc NXS Host, Japan 11413498 epoc NXS Host, Canada 11413506 epoc NXS Host, ROW 11413518 epoc NXS Host, China-Korea 11413583

Z-1215-2023
Recall number
Z-1215-2023
Initiated
December 12, 2022
Classification
Class II
Status
Ongoing
Quantity
4,369 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

pO2 High Recovery in Quality Control When Adjusted for Barometric Pressure

Code information

Host Software Version /Sensor Configuration: 4.10.6/41.1 Product Description Siemens Material Number (SMN) (UDI-DI) epoc NXS Host, US 11413475 00630414606095; epoc NXS Host, EU 11413497 00630414605760; epoc NXS Host, Japan 11413498 00630414605814; epoc NXS Host, Canada 11413506 00630414605821; epoc NXS Host, ROW 11413518 00630414605678; epoc NXS Host, China-Korea 11413583 00630414612447 .

Distribution pattern

Nationwide Foreign: Algeria Argentina Australia Austria Bahrain Belarus Belgium Bolivia Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Fren.Polynesia Georgia Germany Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Ireland Israel Italy Japan Jordan Kazakhstan Kenya Kuwait Laos Latvia Lithuania Luxembourg Malaysia Mali Martinique Mexico Myanmar Netherlands New Caledonia New Zealand Norway Papua Nw Guinea Peru Philippines Poland Portugal Qatar Republic Korea Reunion Romania Russian Fed. San Marino Singapore Slovakia Slovenia Solomon Islands South Africa Spain Sweden Switzerland Taiwan Thailand Tunisia Turkey U.A.E. Ukraine United Kingdom Uruguay Vietnam

Field note

Send feedback