Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91380

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 14, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
W L Gore & Associates, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

GORE CARDIOFORM Septal Occluder, REF: GSX0030A

Z-1048-2023
Recall number
Z-1048-2023
Initiated
December 14, 2022
Classification
Class II
Status
Ongoing
Quantity
14

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Employee error

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Septal Occluders released with incomplete quality test documentation that assesses the release of the device from the delivery system during the index procedure. A potential failure of this quality test could potentially result could result in occluder embolization, anoxic brain injury, central venous or arterial perforation or rupture.

Code information

UDI-DI/Serial Numbers: 00733132631032/25556388, 25556389, 25556390, 25556391, 25556392, 25556393, 25556394, 25556398, 25556399, 25556400, 25556401, 25556402, 25556403, 25556406

Distribution pattern

US Nationwide distribution in the states of NY, CA, FL, CT, TX, AR, IN, MA, MO, SD, VA.

Field note

Send feedback