Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91199

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 18, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Ventana Medical Systems, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001

Z-0907-2023
Recall number
Z-0907-2023
Initiated
November 18, 2022
Classification
Class II
Status
Ongoing
Quantity
8,927 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.

Code information

UDI-DI Code: 04015630967155 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 F30596 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G27114 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G31438 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H03958 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H13049 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H35110 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J12769 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J18775 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J25361 UDI-DI Code: 07613336106709 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 F31427 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G06715 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G16831 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G24564 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H00001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H07164 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H15132 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H18578 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H23050 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J00572 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J01568 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J14244 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J20989 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H08373 UDI-DI Code: 04015630976591 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H17865 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H30521 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H35111 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 J11932 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 J20132

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom, Albania, Armenia, Bangladesh, Belarus, Costa Rica, Cote d'Ivoire, Cuba, Dominican Republic, Egypt, El Salvador, Ghana, Guatemala, Iran, Israel, Jordan, Kenya, Kosovo, Kuwait, Lebanon, Mauritius, Morocco, Myanmar, Nigeria, North Macedonia, Oman, Panama, Philippines, Russian Fed., Saudi Arabia, Serbia, South Africa, Tanzania, Turkey, Turkmenistan and United Arab Emirates. Update (04-Jan-2023): Argentina, Australia, Brazil, Chile, China, Colombia, Ecuador, Hong Kong, Honduras, Indonesia, India, Japan, Korea, Mexico, Malaysia, New Zealand, Peru, Pakistan, Paraguay, Singapore, Thailand, Taiwan, Uruguay and Vietnam.

Field note

Send feedback

We'll only use this to respond to your feedback.