Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 91020

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
October 17, 2022
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Steris Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

GI4000 Electrosurgical Unit - intended to deliver electrosurgical outputs to perform cutting and/or coagulation and coagulation with or without delivery of argon gas, in flexible endoscopic applications. MODEL #(s): G1110001

Z-0241-2023
Recall number
Z-0241-2023
Initiated
October 17, 2022
Classification
Class II
Status
Terminated
Recalling firm
Steris Corporation
Quantity
5 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component design/selection

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result

Code information

UDI: 00724995180256 Serial Numbers: 0032-0822-003, 0032-0622-009, 0032-0922-001, 0032-0822-001, and 0032-0082-002

Distribution pattern

US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia.

Field note

Send feedback

We'll only use this to respond to your feedback.