device · product 1 of 1
Disposable Subdermal Needle Electrode, Pt/lr, 12 x 0.40mm, product number TE/S46/638, 25 needles per package, sterile.
Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.
These labels are deterministic app interpretations, not FDA categories.
Burns of varying degrees can occur to patients with electrodes left in while in the MRI scanner as a result of inadequate safety information in the IFU.
Code information
Lot numbers 044220, exp. 1/1/2023; 045077, exp. 3/1/2023; 045443, exp. 4/1/2023; and 049285, exp. 7/1/2024; UDI-DI 08718375861530.
Distribution pattern
US Nationwide distribution in the states of CA, FL, NH, OH, and WA.