Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 90195

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 20, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Olympus Corporation of the Americas

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Olympus CYSTONEPHROFIBERSCOPE Model: CYF-5

Z-1262-2022
Recall number
Z-1262-2022
Initiated
April 20, 2022
Classification
Class II
Status
Ongoing
Quantity
7 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Code information

UDI-DI: 04953170339417 Serial Number: 2456630 7171632 2769732 W501648A 2355088 2557155 2972060

Distribution pattern

CA FL MA MD MI NJ OH PR WI

device · product 2 of 3

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6

Z-1263-2022
Recall number
Z-1263-2022
Initiated
April 20, 2022
Classification
Class II
Status
Ongoing
Quantity
2 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Code information

UDI-DI: 04953170340802 Serial Number: 2836210 2834747I

Distribution pattern

CA FL MA MD MI NJ OH PR WI

device · product 3 of 3

Olympus URETERO-RENO FIBERSCOPE Model: URF-P6R

Z-1264-2022
Recall number
Z-1264-2022
Initiated
April 20, 2022
Classification
Class II
Status
Ongoing
Quantity
4 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Incorrect adhesive used to secure the diopter ring and nameplate ring. Improperly connected diopter ring may result in the loss of adjustment function leading to a blurred field of vision and unintentional contact with the patients body and in rare cases perforation in the urinary tract/renal pelvis.

Code information

UDI-DI: 04953170340833 Serial Number: 2836209F 2835978 2837272 2836299

Distribution pattern

CA FL MA MD MI NJ OH PR WI

Field note

Send feedback

We'll only use this to respond to your feedback.