Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89699

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 22, 2022
Product types
Device
Classifications
Class II
Statuses
Completed
Recalling firm wording
Illumina, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Illumina NextSeq 550Dx, REF: 20005715, CE, IVD

Z-0854-2022
Recall number
Z-0854-2022
Initiated
February 22, 2022
Classification
Class II
Status
Completed
Recalling firm
Illumina, Inc.
Quantity
674 instruments

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.

Code information

Lot #s NDX550110 - NDX550843 ; UDI: (01)00816270020125

Distribution pattern

US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN OUS: Australia Austria Belgium Chile China Denmark Finland France Germany Hungary Ireland Italy Japan Latvia Netherlands Norway Poland Portugal Romania Russian Federation Saudi Arabia Singapore Slovakia South Africa South Korea Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom Vietnam

Field note

Send feedback