device · product 1 of 1
SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x
- Recall number
- Z-0885-2022
- Initiated
- January 19, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Volcano Corporation
- Quantity
- 3,216
App-derived interpretation
If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
Code information
All serial numbers, limited to cases when: The SyncVision runs the current supported software version 4.2.x, AND The secondary modality (IntraSight IVUS) runs software version 5.x. Device Identifier (DI): 00845225012434
Distribution pattern
US: WI, CA, CO, MO, ID, DC, LA, VA, KS, OK, GA, KY, IL, AZ, NY, IN, FL, TX, NC, OH, WA, OR, NM, MT, IA, MI, TN, MN, PA, AL, MA, WV, NV, NH, CT, NE, NJ, HI, AR, UT, MD, SC, MS, AK, DE OUS: JP, SA, GB, DE, PT, SG, PL, ZA, NL, MT, ES, IN, GR, EG, FI, HK, FR, NO, DK, BE, AT, IT, CO, KW, MX, PS, QA, CY, CA, TW, LV, SE, HR, SI, MY, KR, IL, TH, AU, BR, CL, SK, CH, LB, OM, ID, PK, IE, NZ, PR, RO, MM, AR, RS, RU, PA, KE, UZ