device · product 1 of 5
KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299
- Recall number
- Z-0585-2022
- Initiated
- February 18, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Microbiologics Inc
- Quantity
- 1,975 units
App-derived interpretation
out of specification
Official device-enrichment evidence · Sourced
Process design
Inspect official wording and provenance
Reason for recall
Tests returning incorrect susceptible results. High Level Gentamicin Synergy value was out of specification (sensitive) when analyzed on the Vitek 2 AST-GP67 card, and these lots failed the disk diffusion method used for product release.
Code information
a. Catalog Number 0959P, UDI: 20845357023210, Lot Number (Expiration Date): 959-94-3 (04/30/2020), 959-95-2 (09/30/2020), 959-96-1 (12/31/2020), 959-96-4 (12/31/2021), 959-97-4 (03/31/2021), 959-97-5 (03/31/2021); b. Catalog Number 0959K, UDI: 30845357023224, Lot Number (Expiration Date): 959-94-2 (04/30/2020), 959-95-1 (09/30/2020), 959-95-5 (09/30/2020), 959-96-3 (12/31/2020), 959-97-1 (03/31/2021), 959-97-6 (03/31/2021);
Distribution pattern
Worldwide Distribution. US nationwide including Puerto Rico and Guam, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia/Hrvatska, Curacao, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Ireland, Israel, Italy, Japan, Democratic People's Republic of Korea, Republic of Korea, Kuwait, Lebanon, Malaysia, Netherlands, New Zealand, Panama, Peru, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.