Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89283

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 22, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Covidien, LP

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

SonarMed AirWave Monitor Kit (Monitor, Sensor Cable, Power Cord), Item Code AW-M0001. used with endotracheal tube (ETT)

Z-0514-2022
Recall number
Z-0514-2022
Initiated
December 22, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Covidien, LP
Quantity
66 devices

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Prior to use on a patient, the device alarm speaker failed to annunciate on power up and the device falsely detected a sensor cable disconnect and reconnect during setup. The monitor detection of a false sensor cable disconnect and reconnect may cause the monitor to discontinue monitoring and transition into the calibration screen without an audible alarm.

Code information

UDI-Device Identifier (GTIN/UPN): 00851334007001; Serial Numbers: AW0501 AW0519 AW0537 AW0556 AW0502 AW0520 AW0540 AW0557 AW0503 AW0521 AW0541 AW0558 AW0504 AW0522 AW0542 AW0559 AW0505 AW0524 AW0543 AW0560 AW0506 AW0525 AW0544 AW0561 AW0507 AW0526 AW0545 AW0562 AW0508 AW0527 AW0546 AW0563 AW0509 AW0528 AW0548 AW0564 AW0510 AW0529 AW0549 AW0565 AW0511 AW0530 AW0550 AW0566 AW0512 AW0531 AW0551 AW0567 AW0513 AW0532 AW0552 AW0568 AW0514 AW0533 AW0553 AW0569 AW0515 AW0536 AW0554 AW0570 AW0516 AW0538 AW0555 AW0571 AW0517 AW0539

Distribution pattern

The devices were distributed to the following US states: CA, CO, FL, IL, KY, MA, OH, TN, and TX.

Field note

Send feedback