basixALPHA Inflation Device 30 atm/bar 20 mL - Product Usage: Used to inflate and deflate an angioplasty balloon or other interventional device, and to measure the pressure within the balloon.
- Recall number
- Z-2166-2021
- Initiated
- June 10, 2021
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 7,211 devices
App-derived interpretation
design issue
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Due to a design issue, inflation syringe handle separating from the inflation device when handle is withdrawn.
Code information
Catalog Number: IN3130 UDI: 00884450440179; 00884450510155; Lot Numbers: H1862928; H1896261; H1923761; H1929106; H1939255; H1966485; H1978295; H2068264; H2073193; H2073247; H2077173; Catalog Number: IN3152 UDI: 00884450511220 Lot Numbers: H1896262; H1909581; H1923762; H1929107; H1978299; H2007415; H2031599; H2068265; H2073248; H2077175; H2084368; Catalog Number: IN3302 UDI: 00884450511305 Lot Numbers: H1896263; H1966489; Catalog Number: IN3802 UDI: 00884450511510 Lot Numbers: H18962650; H1939259; H1966490; H1973390; H1978305; H1993118; H2001426; H2007417; H2012534; H2031601; H2073198; H2073249; H2077176;
Distribution pattern
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, ID, IL IN, KS, LA, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, WV, and WY. The countries of Australia, Austria, China, Denmark, Finland, France, Germany, Israel, Japan, Kenya, Kyrgyzstan, Netherlands, New Zealand, Norway, Qatar, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates and United Kingdom.