Event 87882

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 30, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Nico Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

NICO Myriad-LX Illumination Pack: REF: NN-1000 Mvriad-LX Illumination Pack. 11ga REF: NN-1001 Mvriad-LX Illumination Pack. l3ga Intended to perform resection and removal of soft tissues and fluids under direct visualization accessories to the NICO Myriad System.

Z-1839-2021
Recall number
Z-1839-2021
Initiated
April 30, 2021
Classification
Class II
Status
Terminated
Recalling firm
Nico Corp.
Quantity
439 packs

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or end user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or end user.

Code information

Lot Numbers: 37KII300 37KI1301 37LI0320 37LI0322 37AJ1438 37AJ1439 37BJ0513 37BJ0516 37CJ0203 37CJ0204 37GJ1603 37GJ1607 37110208 37IJ0209

Distribution pattern

US Nationwide Distribution

device · product 2 of 2

Myriad-LX Light Source-for delivery of light to the surgical field to enhance visualization of tissue during resection Part Number NN-7013 The Myriad-LX light source is a high-intensity 300W Xenon illuminator with a touch screen and digital dimming controls. The source is fitted with a proprietary connector to ensure usage is reserved for the Myriad-LX Illumination Fiber.

Z-1840-2021
Recall number
Z-1840-2021
Initiated
April 30, 2021
Classification
Class II
Status
Terminated
Recalling firm
Nico Corp.
Quantity
80 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or end user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient or end user.

Code information

Serial Numbers 350-00194D 350-00355D 350-00356D 350-00357D 350-00376D through 350-00444D 350-00451D through 350-00460D 350-00463D through 350-00467D 350-004710 through 350-00475D

Distribution pattern

US Nationwide Distribution