Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Catalogue Number: CAL2351
- Recall number
- Z-1563-2021
- Initiated
- March 29, 2021
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Randox Laboratories Ltd.
- Quantity
- 98 kits
App-derived interpretation
Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.
Code information
Batch / Lot number: 1162UE GTIN: 05055273200966
Distribution pattern
Domestic: IN, MA, CA, NJ, TX, NE, NY, VA, MD, MN, OR, IL, CT, DC, MS. Foreign: Albania, Argentina, Australia, Bangladesh, Bolivia, Bosnia & Herzegovina, Cameroon, Canada, Colombia, Cyprus, Denmark, Egypt, Estonia, Fiji, France, Germany, Greece, Hong Kong, Iraq, Italy, Ireland, Jordan, Kazakhstan, Kenya, Lithuania, Luxembourg, Mauritius, Mexico, Myanmar, Nepal, New Zealand, Pakistan, Palestine, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Thailand, The Netherlands, Turkey, UK, Ukraine, USA, and Vietnam.