Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87440

3 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 26, 2021
Product types
Device
Classifications
Class II
Statuses
Completed
Recalling firm wording
Biosense Webster, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

3 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 3

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (SMALL) C3, REF: D138501, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016253 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Z-1285-2021
Recall number
Z-1285-2021
Initiated
February 26, 2021
Classification
Class II
Status
Completed
Recalling firm
Biosense Webster, Inc.
Quantity
71,473 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Code information

All Lots

Distribution pattern

Worldwide distribution.

device · product 2 of 3

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (Large) C3, REF: D138503, Rx Only, CE 2797, STERILE EO, udi: (01)10846835016260 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Z-1286-2021
Recall number
Z-1286-2021
Initiated
February 26, 2021
Classification
Class II
Status
Completed
Recalling firm
Biosense Webster, Inc.
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Code information

All Lots

Distribution pattern

Worldwide distribution.

device · product 3 of 3

Biosense Webster, CARTO VIZIGO, Bi-Directional Guiding Sheath, 8.5F (medium) C3, REF: D138502, Rx Only, CE 2797, STERILE EO, UDI: (01)10846835016277 - Product Usage: provide accessibility and maneuverability in the cardiac anatomy during cardiac electrophysiological mapping and ablation procedures.

Z-1287-2021
Recall number
Z-1287-2021
Initiated
February 26, 2021
Classification
Class II
Status
Completed
Recalling firm
Biosense Webster, Inc.
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism

Code information

All Lots

Distribution pattern

Worldwide distribution.