Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87263

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 29, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Confirm Biosciences Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Healgen COVID-19 IgG/IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) REF GCC)V-402a - Product Usage: is a lateral flow immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV- 2 in human venous whole blood, plasma from anticoagulated blood or serum.

Z-1125-2021
Recall number
Z-1125-2021
Initiated
January 29, 2021
Classification
Class II
Status
Terminated
Recalling firm
Confirm Biosciences Inc
Quantity
563,840 cassettes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).

Code information

Catalog Number: HE Cassette COVID 19EUA Lot Numbers: 2006167eua 2006168eua 2006169eua

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. The countries of Australia, Mexico, Honduras, Spain, Sierra Leone, Bahamas, Ecuador, Turkey, Romania, Kuwait, Libya, Qatar, Ghana, Argentina, Bolivia, Brazil, Guatemala, Panama, India, United Kingdom, West Indies, South Africa, United Arab Emirates, Slovenia, Canada, Israel, United Kingdom, Romania, Argentina, Honduras, Russian Federation, R¿publique D¿mocratique du Congo, Switzerland, Dominican Republic, Greece, Saudi Arabia, and Portugal.

device · product 2 of 2

Orient Gene COVID- 19 IgG/ IgM Rapid Test Cassette (Whole Blood/Serum/Plasma) REF GCCOV-402a - Product Usage: is a lateral flow immunoassay intended for the qualitative detection and differentiation of IgM and IgG antibodies to SARS-CoV- 2 in human venous whole blood, plasma from anticoagulated blood or serum.

Z-1126-2021
Recall number
Z-1126-2021
Initiated
January 29, 2021
Classification
Class II
Status
Terminated
Recalling firm
Confirm Biosciences Inc
Quantity
2,003,053 cassettes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Other

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to two (2) issues: 1) The firm distributed COVID-19 Rapid Test Cassettes outside of the Emergency Use Authorization to non-certified Clinical Laboratories. 2) COVID-19 Rapid Test Cassettes were not adequately validated for its intended use (fingerstick tests for in-home use/testing).

Code information

Catalog Number: HE Cassette COVID 19 Lot Numbers: 2002198, 2003244, 2003285, 2003290, 2003291, 2003295, 2003308, 2004148, 2004158, 2005153, 2005154, 2005155, 2005213

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. The countries of Australia, Mexico, Honduras, Spain, Sierra Leone, Bahamas, Ecuador, Turkey, Romania, Kuwait, Libya, Qatar, Ghana, Argentina, Bolivia, Brazil, Guatemala, Panama, India, United Kingdom, West Indies, South Africa, United Arab Emirates, Slovenia, Canada, Israel, United Kingdom, Romania, Argentina, Honduras, Russian Federation, R¿publique D¿mocratique du Congo, Switzerland, Dominican Republic, Greece, Saudi Arabia, and Portugal.