Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86848

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 18, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips North America Llc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Philips IntelliVue X2 Multi measurement Server: M3002A (865039) - Product Usage: intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside and outside of the hospital environment.

Z-0544-2021
Recall number
Z-0544-2021
Initiated
November 18, 2020
Classification
Class II
Status
Terminated
Recalling firm
Philips North America Llc
Quantity
16 total: 2 units US ; 14 OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.

Code information

Serial Numbers: US: DE73607502 DE73607498 OUS: System Serial Number DE95038935 DE73608191 DE73608209 DE73601727 DE73604240 DE73601547 DE73601596 DE73604182 DE83617535 DE73608259 DE83617500 DE83617522 DE73601651

Distribution pattern

Worldwide distribution - US Nationwide distribution in the state of SD and the countries of China, England, Germany, Hong Kong, Portugal.

device · product 2 of 2

Philips IntelliVue MP2 Patient Monitors (865040) - Product Usage: intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults, pediatrics, and neonates in a hospital environment and during patient transport inside and outside of the hospital environment.

Z-0545-2021
Recall number
Z-0545-2021
Initiated
November 18, 2020
Classification
Class II
Status
Terminated
Recalling firm
Philips North America Llc
Quantity
2 units OUS

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.

Code information

Serial Numbers: DE73602213 DE73602577 UDI: (01)00884838000278(21)DE73602213 (01)00884838000278(21)DE73602577

Distribution pattern

Worldwide distribution - US Nationwide distribution in the state of SD and the countries of China, England, Germany, Hong Kong, Portugal.

Field note

Send feedback