Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86366

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 06, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Carl Zeiss Meditec AG

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.

Z-0613-2021
Recall number
Z-0613-2021
Initiated
August 06, 2020
Classification
Class II
Status
Terminated
Recalling firm
Carl Zeiss Meditec AG
Quantity
603 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Code information

Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.

device · product 2 of 2

Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses.

Z-0614-2021
Recall number
Z-0614-2021
Initiated
August 06, 2020
Classification
Class II
Status
Terminated
Recalling firm
Carl Zeiss Meditec AG
Quantity
132 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Code information

Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TX, UT, VA, WA and WY. The countries of Argentina, Austria, Australia, Belgium, Bolivia, Canada, Colombia, Croatia, Cyprus, Denmark, France, Germany, Hong Kong, Hungary, India, Ireland, Italy, Japan, Latvia, Luxembourg, Malaysia, Maldives, Netherlands, Portugal, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Turkey, and United Kingdom.

Field note