Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 86363

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 28, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Greiner Bio-One North America, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VACUETTE TUBE 5 mL CAT Serum Separator Clot Activator, Item Number: 456018P - Product Usage: are used to collect, transport, and process blood for testing serum, plasma or whole blood in the clinical laboratory.

Z-3027-2020
Recall number
Z-3027-2020
Initiated
August 28, 2020
Classification
Class II
Status
Terminated
Quantity
1,008,000 pieces

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Blood collection tubes experienced an issue with separation and clotting

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Blood collection tubes experienced an issue with separation and clotting

Code information

Lot Number: B2003399

Distribution pattern

US Nationwide distribution including in the states of IL, PA, VA, MO, NY.

Field note

Send feedback

We'll only use this to respond to your feedback.