device · product 1 of 2
Renuvion/ J-Plasma Precise Open Handpiece. Catalog numbers BVX-044-BPS, BVX-044-BPP, BVX-150-BPP, BVX-150-BPS, APYX-044-BPS, APYX-150-BPP, and APYX-150-BPS - Product Usage: for electrosurgical cutting, coagulation and ablation of soft tissue during open surgical procedures.
- Recall number
- Z-2060-2020
- Initiated
- April 20, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Bovie Medical Corporation
- Quantity
- 13,948
App-derived interpretation
Unexpected stress fractures on the shaft of the hand piece may result in fragmentation during clinical use.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Unexpected stress fractures on the shaft of the hand piece may result in fragmentation during clinical use.
Code information
The following Catalog numbers which were distributed from 06/01/2016 through present: BVX-044-BPS, BVX-044-BPP, BVX-150-BPP, BVX-150-BPS, APYX-044-BPS, APYX-150-BPP, APYX-150-BPS
Distribution pattern
Worldwide distribution - US nationwide including in the states of Puerto Rico, and the countries of CA, JO, CO, MX, BR, CR, PA, KW, CL, MA, ES, TH, PH, IQ, LB, DO, Japan, QA, AE, PE, AU, VN, IL, Lithuania, UAE, Belgium, Denmark, Austria, Poland, Germany, Slovakia, Spain, Cyprus, Switzerland, Turkey, UK, Italy, Bulgaria, Greece, Norway, Ireland, Hungary, and Croatia.