Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 85535

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 20, 2020
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Trividia Health, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

TRUE METRIX SELF MONITORING BLOOD GLUCOSE SYSTEM - Product Usage: is intended for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip or forearm.

Z-2201-2020
Recall number
Z-2201-2020
Initiated
April 20, 2020
Classification
Class II
Status
Terminated
Recalling firm
Trividia Health, Inc.
Quantity
1 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

One (1) TRUE METRIX AIR blood glucose meter distributed in the United States was packaged into a TRUE METRIX blood glucose meter kit and has an incorrect factory-set unit of measure; the meter displays glucose results in mmol/L rather than mg/dL.

Code information

Serial Number TA1548753

Distribution pattern

US Nationwide distribution including in the state of Florida.

Field note

Send feedback

We'll only use this to respond to your feedback.