device · product 1 of 1
inLight Medical LED Pads used with Polychromatic light technology system
- Recall number
- Z-1943-2021
- Initiated
- June 04, 2019
- Classification
- Class III
- Status
- Terminated
- Recalling firm
- Inlightened Partners LLC
- Quantity
- 926
App-derived interpretation
LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.
Official device-enrichment evidence · Sourced
Component change control
Inspect official wording and provenance
Reason for recall
LED pads with cord connectors that have v-shaped notches may not have been assembled properly. This issue could lead to a power interruption, causing the controller and power supply to shut down.
Code information
All serial numbers with cord connectors that have v-shaped notches
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of CA, TX, NC, ID, CO, VA, FL, VT, AK, DC, MA, IL, AZ, MN, IA, TN, WI, WA, OK, MI, OH, GA, NY, IN, PA, NV, NJ, AR, CO , LA, MT, OR, SD, ND, MD, UT, MO, KY, NM, KS, AS, NE, WY, PR, MS and the country of Canada.