device · product 1 of 1
Atellica CH Ethyl Alcohol, Test Code ETOH, Siemens Material Number (SMN) 11097501, UDI 00630414596112. The product is an ethyl alcohol assay, IVD use.
- Recall number
- Z-1599-2020
- Initiated
- February 14, 2020
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 1124
App-derived interpretation
The assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
The assay may exhibit a positive bias in QC and patient results within the 10-day pack calibration interval.
Code information
All in-date lots: 190133, 190222 ***Updated 4/6/2020*** New lot added: 100066 ***Updated 9/4/20*** New lot added: 100168
Distribution pattern
Domestic distribution nationwide. Foreign distribution to Australia, Austria, Bahamas, Belgium, Canada, Curacao, Chile, Colombia, Czech Republic, Denmark, Finland, France, Germany, India, Israel, Italy, Japan, Kenya, Latvia, Mexico, Netherlands, Norway, Poland, Portugal, Slovakia, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, U.A.E., Uruguay, United Kingdom, and Vietnam.