Event summary

Timeline bucket
August 30, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Ultrasound Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Transducer S7-3t UDI (01)00884838061439 REF 989605406772 Can be used with: CX50, EPIQ 7, Affiniti 50 and Affiniti 70 Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Z-1841-2020
Recall number
Z-1841-2020
Initiated
August 30, 2019
Classification
Class II
Status
Terminated
Recalling firm
Philips Ultrasound Inc
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Code information

All lots.

Distribution pattern

Worldwide distribution - US Nationwide distribution.

device · product 2 of 4

Transducer S7-3t UDI (01)00884838067530 REF 989605439232 Used as accessory with: iE33, HD11 Ultrasound Systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Z-1842-2020
Recall number
Z-1842-2020
Initiated
August 30, 2019
Classification
Class II
Status
Terminated
Recalling firm
Philips Ultrasound Inc
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Code information

All lots.

Distribution pattern

Worldwide distribution - US Nationwide distribution.

device · product 3 of 4

Transducer S8-3t UDI (01)00884838073524 REF 989605431171 Can be used as accessory with: EPIQ 7, Affiniti 70, EPIQ 5, EPIQ CVxi Ultrasound systems - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Z-1843-2020
Recall number
Z-1843-2020
Initiated
August 30, 2019
Classification
Class II
Status
Terminated
Recalling firm
Philips Ultrasound Inc
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Code information

All lots.

Distribution pattern

Worldwide distribution - US Nationwide distribution.

device · product 4 of 4

Transducer S8-3t UDI (01)00884838067523 REF 989605420183 Can be used with: iE33 Utrasound System - Product Usage: Diagnostic ultrasound imaging or fluid flow analysis of the human body - particularly transesophageal echocardiography (TEE) studies

Z-1844-2020
Recall number
Z-1844-2020
Initiated
August 30, 2019
Classification
Class II
Status
Terminated
Recalling firm
Philips Ultrasound Inc
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of the transducer. This issue may result in affected transducers not initiating auto-cool during use which could result in patient injury

Code information

All lots.

Distribution pattern

Worldwide distribution - US Nationwide distribution.