Event summary

Timeline bucket
December 19, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Randox Laboratories, Limited

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

RX Daytona Plus (with ISE/without ISE); Models Nos. RX4040 (w/ISE) and RX4041 (without ISE) - Product Usage: A fully automated, random access, clinical chemistry analyzer complete with dedicated analyzer software. For use in clinical laboratories.

Z-1349-2020
Recall number
Z-1349-2020
Initiated
December 19, 2019
Classification
Class II
Status
Terminated
Quantity
U.S.: RX4040-2 units RX 4041-3 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect color flagging of IQC as "out of range."

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software version UI2550642107 for the RX Daytona + instrument released to correct the following issues: 1) The order of samples displayed is not sequential according to SID (sample identification number) when searching the run/results screen; 2) Incorrect color flagging of IQC as "out of range."

Code information

Software version UI2550642107 for the RX Daytona Plus instrument.

Distribution pattern

US Nationwide distribution including state of PR.