openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Siemens ARTIS pheno, Model # 10849000. Interventional Fluoroscopic X-Ray System - Product Usage: Angiography system for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.
These labels are deterministic app interpretations, not FDA categories.
The potential exists for system movement to be permanently blocked by an activation of the collision sensor integrated in the cover of the collimator even when no collision has occurred.
Code information
Serial Numbers 164231, 164234, 164240, 164241, 164242, 164244
Distribution pattern
US Nationwide distribution in the states of FL, IL, IN, MA, PA, VA.