Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84547

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 13, 2019
Product types
Device
Classifications
Class III
Statuses
Terminated
Recalling firm wording
Randox Laboratories Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Liquid Clinical Chemistry Control, Catalogue Number: LAE4217, Size: 12 x 5 ml

Z-1100-2020
Recall number
Z-1100-2020
Initiated
December 13, 2019
Classification
Class III
Status
Terminated
Recalling firm
Randox Laboratories Ltd.
Quantity
2

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Error in labeling

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86).

Code information

Catalogue Number: LAE4217 GTIN: 05055273208993 Lot Number: 1021UE (EXP: 2020-06-28)

Distribution pattern

US: NY and DE OUS: France, Lithuania, Poland, Spain, Switzerland, Turkey, Australia, Hong Kong, Bosnia and Herzegovina, Costa Rica, Ecuador, Gambia, New Zealand and Solomon Islands.

Field note

Send feedback