openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 2
BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
These labels are deterministic app interpretations, not FDA categories.
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
Code information
Serial Numbers: W1-6C000006, W1-71000007, W1-73000009, W1-73000010, XP-68000687T, W1-71000008, W1-85000011
Distribution pattern
The products were distributed to the following US states: MI and TX.
device · product 2 of 2
ABS800 Chemistry Analyzer Part Number BA81F-PA00001, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
These labels are deterministic app interpretations, not FDA categories.
The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
Code information
Serial Numbers W2-6C000001, W2-6C000002, W2-6C000003, W2-6C000004, W2-6C000005, XP-5B000611T
Distribution pattern
The products were distributed to the following US states: MI and TX.