Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84399

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 15, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Arrow International Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Arrow Berman Angiographic Catheter 50cc Item Number: AI-07130

Z-0715-2020
Recall number
Z-0715-2020
Initiated
November 15, 2019
Classification
Class II
Status
Terminated
Recalling firm
Arrow International Inc
Quantity
71 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

Code information

Lot Number: 16F19H0089

Distribution pattern

U.S. Nationwide, Canada

device · product 2 of 2

Arrow Berman Angiographic Catheter 60cm Item Number: AI-07130-J

Z-0716-2020
Recall number
Z-0716-2020
Initiated
November 15, 2019
Classification
Class II
Status
Terminated
Recalling firm
Arrow International Inc
Quantity
1 unit

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Potential leak in the balloon extension line. If a leak was present and went undetected during pre-use testing as outlined in the IFU, this would cause the balloon to deflate slowly following placement. If the leak went undetected by attending clinicians during use, it could result in contrast injection into the vascular or cardiac tissue which could result in serious injury or death

Code information

Lot Number: 16F19H0093

Distribution pattern

U.S. Nationwide, Canada

Field note

Send feedback