Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 84135

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 08, 2019
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Canon Medical System, USA, INC.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Z-0360-2020
Recall number
Z-0360-2020
Initiated
February 08, 2019
Classification
Class II
Status
Terminated
Quantity
9 systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Error in labeling

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Code information

Model CAT-870B D053991 Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.

Field note

Send feedback